Journal of Physics: Conference Series, vol: 694,1 (2016)
Using computer aided system to determine the maximum depth of visualization of B-Mode diagnostic ultrasound image
Maslebu G., Adi K., Suryono
Abstract
In the service unit of radiology, ultrasound modality is widely used because it has advantages over other modalities, such as relatively inexpensive, non-invasive, does not use ionizing radiation, and portable. Until now, the method for determining the depth visualization on quality control program is through the visual observation of ultrasound image on the monitor. The purpose of this study is to develop a computer-aided system to determine maximum depth of visualization. Data acquisition was done by using B-Mode Diagnostic Ultrasound machine and Multi-purpose Multi-tissue Ultrasound Phantom model 040GSE. Phantom was scanned at fixed frequency of 1,8 MHz, 2,2 MHz, 3,6 MHz and 5,0 MHz with a gain variation of 30 dB, 45 dB, and 60 dB. Global thresholding and Euclidean distance method were used to determine maximum visualization depth. From this study, it is proved that the visualization depth using computer aided provide deeper visualization than visual interpretation. The differences between expert verification and the result of image processing are <6%. Thus, computer aided system can be used for the purpose of quality control in determining maximum visualization depth of B-Mode diagnostic ultrasound image. © Published under licence by IOP Publishing Ltd.
Keyword: